Skip to main navigation Skip to search Skip to main content

Repercusión de la actualización del software del monitor 5008 en el volumen convectivo total

Translated title of the contribution: Impact of the 5008 monitor software update on total convective volume
  • Francisco Maduell*
  • , Néstor Rodríguez
  • , Laura Sahdalá
  • , Diego Coronel
  • , Marta Arias-Guillén
  • , Raquel Ojeda
  • , Manel Vera
  • , Néstor Fontseré
  • , Aleix Cases
  • , Josep M. Campistol
  • *Corresponding author for this work
  • Servicio de Nefrología, Hospital Clínic

Research output: Contribution to journalArticlepeer-review

17 Scopus citations

Abstract

Introduction: On-line haemodiafiltration (OL-HDF) is based on the controlled administration of large volumes of replacement ultrapure dialysate to the extracorporeal blood circuit of dialysis. It is currently the most effective technique for eliminating small and large uraemic toxins. Recent studies have observed a direct association between a decrease in mortality and the received convective volume. The latest Fresenius 5008 (5008 CorDiax) monitor software update enables the automation of the replacement volume without requiring total protein and haematocrit values, with the aim of maximising convection. Objective: The study's objective was to evaluate the recent version of the 5008 monitor software compared to the previous version on the impact on total convective volume. Material and method: We included 63 patients (44 males and 19 females), with a mean age of 65.2 ± 15 years, on OL-HDF. Each patient was analysed in 6 sessions; 3 with the 5008 monitor and 3 with the 5008 CorDiax monitor. The replacement volume, total convective volume and dialysis parameters were determined in each session. Results: No significant differences were observed in blood, venous or transmembrane pressure, nor was there an increase in the number of alarms or coagulation of lines or dialysers. A significant increase of replacement fluid volume, from 27.2 to 31.2 l/session, was observed using the CorDiax software. Total convective volume increased from 29.5 to 33.3 l/session, representing an effective convective volume increase from 26 % to 29.6 % of total filtered blood. Conclusion: The 5008 dialysis monitor software update has meant a 13 % increase of total convective volume and a 3.5 % increase of total filtered blood.

Translated title of the contributionImpact of the 5008 monitor software update on total convective volume
Original languageSpanish
Pages (from-to)599-604
Number of pages6
JournalNefrologia
Volume34
Issue number5
DOIs
StatePublished - 2014
Externally publishedYes