New QRS detection algorithm based on the Hilbert transform

D. S. Benitez, P. A. Gaydecki, A. Zaidi, A. P. Fitzpatrick

Producción científica: Contribución a una revistaArtículo de la conferenciarevisión exhaustiva

168 Citas (Scopus)

Resumen

A robust new algorithm for QRS detection using the properties of the Hilbert transform is proposed in this paper. The method allows R waves to be differentiated from large, peaked T and P waves with a high degree of accuracy and minimizes the problems associated with baseline drift, motion artifacts and muscular noise. The performance of the algorithm was tested using the records of the MIT-BIH Arrhythmia Database. Beat by beat comparison was performed according to the recommendation of the American National Standard for ambulatory ECG analyzers (ANSI/AAMI EC38-1998). A QRS detection rate of 99.64%, a sensitivity of 99.81% and a positive prediction of 99.83% was achieved against the MIT-BIH Arrhythmia database. The noise tolerance of the new proposed QRS detector was also tested using standard records from the MIT-BIH Noise Stress Test Database. The sensitivity of the detector remains about 94% even for signal-to-noise ratios (SNR) as low as 6dB.

Idioma originalInglés
Páginas (desde-hasta)379-382
Número de páginas4
PublicaciónComputers in Cardiology
EstadoPublicada - 2000
Publicado de forma externa